Search
Hydrogen is the future, we can sense it!™
 
 



HY-OPTIMA™ 1700 Intrinsically Safe Process Hydrogen Monitor
 
The H2scan 3rd Generation Handheld Hydrogen Sensing Unit - Click To Enlarge
*Pictures shown may include future optional equipment

Online Customer Application
H2scan’s HY-OPTIMATM 1700 in-line, real time hydrogen-specific process analyzers are designed for ease of use, interface flexibility and true process control. The HY-OPTIMA™ 1700 is an intrinsically safe device that incorporates our patented solid-state sensor and is configured to operate in background gas streams with up to 95% relative humidity and temperatures up to100ºC. The HY-OPTIMA™ 1700 is ideal for process gas streams containing virtually no contaminants and where real time continuous measurements of hydrogen can increase process plant efficiencies, improve diagnostics and enhance maintenance management translating into higher profitability.

 

Performance

Hydrogen Sensitivity Range:
0.5% to 100% hydrogen by volume at 1 ATM
Response Time: T90 < 30 sec
Ingress Protection: IP64 capable
Recommended Verification Interval: 90 days
Product Life Expectancy: 10 years
Accuracy(*):
± 0.3% absolute for 0.5 to 10% H2
± 1.0% absolute for 10 to 100% H2
Drift/week:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Repeatability:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Linearity:
± 0.2% absolute for 0.5-10% H2
± 0.4% absolute for 10-100% H2


Interface Options

Power Barrier Input Voltage: 24 VDC nominal
Power Barrier Input Voltage Range: 20 VDC to 28 VDC
Intrinsically Safe Sensor Input Voltage: 10 VDC
Sensor Input Voltage Range: 5 VDC to 28 VDC
Input Power: 10 W
Analog Output: 4 mA to 20 mA
Serial Communication: RS422


Operating Conditions

Operating Humidity: 0% to 95% RH
Flow Rate: 0.1 slpm to 50 slpm
Pressure: 0 to 7 bar gauge, 0 to 100 psi gauge
Process Gas Temperature Range: -20ºC to 100ºC
Operating Temperature Range: -20ºC to 40ºC
Storage Temperature Range: -40ºC to 50ºC
Calibration Background Gas: Nitrogen


Included Accessories 

Intrinsically Safe Power Barrier
4 m Power / Analog Cable
Adapter Fitting
• ½ in. MNPT thread
• ½ in. FNPT thread
• -8 SAE/MS thread size
• many other industry standards


Optional Accessories 

Single Channel Intrinsically Safe Analog Output Barrier (with or without relays)
Single Channel Intrinsically Safe RS422 Serial Interface Barrier
Serial Interface Cable: 4 m (custom lengths available)
Armored Power / Analog Cable (custom lengths available)


Dimensions

Total Length: 10.3 in.
Length of Enclosure: 6.8 in.
Length of Tube: 2.9 in.
Width: 3.4 in.
Depth: 1.4 in.







 
 